Welcome ACMAN factory,come to understand our strength.
Dust Extraction Systems for Air Pollution Control
Healthy Manufacture
Home / All / BIBO Filter Housing Unit /

F9+H14 Safe Change Bag In Bag Out (BIBO) Type with Housing In Low Level Return Air Riser Arrangement

F9+H14 Safe Change Bag In Bag Out (BIBO) Type with Housing In Low Level Return Air Riser Arrangement

F9+H14 Safe Change Bag In Bag Out (BIBO) Type with Housing In Low Level Return Air  Riser Arrangement
F9+H14 Safe Change Bag In Bag Out (BIBO) Type with Housing In Low Level Return Air  Riser ArrangementF9+H14 Safe Change Bag In Bag Out (BIBO) Type with Housing In Low Level Return Air  Riser ArrangementF9+H14 Safe Change Bag In Bag Out (BIBO) Type with Housing In Low Level Return Air  Riser ArrangementF9+H14 Safe Change Bag In Bag Out (BIBO) Type with Housing In Low Level Return Air  Riser Arrangement
CategoriesBIBO Filter Housing Unit
BrandACMAN
ModelBIBO-FH-CFM
FOB portSHANGHAI/NINGBO
Terms of PaymentT/T
Update TimeMar 29,2024
Detail Information
F9+H14 Two Stage Safe Change Bag In Bag Out (BIBO) Type with Housing In Low Level Return Air  Riser Arrangement
Bag-in and bag-out filters appeared in the 1950s and 1960s, and are currently widely used in a variety of high-risk or high isolation requirements, such as: chemical and biological (CB) protection, chemical, biological and radiological ( CBR) protection, nuclear chemical, biological and chemical (NBC) protection, hospital isolation room, pharmaceutical equipment, microelectronics environment, food processing area, biological research, genetic and biotechnology laboratory, industrial processing discharge system, chemical processing equipment, animal Disease research laboratories, radioisotope processing equipment, nuclear power plants, strategic nuclear facilities, DOE equipment, military bases, etc.
How does two stage F9+H14 BIBO housing working?
The equipment design refers to the leakage limit of the international OEB standard, and the overall design is in accordance with the control of the 1ug/m³ dust leakage rate and meets the requirements of the OEB5 standard

  Bag-in-bag-out filter, namely Bag In Bag Out Filter, usually referred to as BIBO, is also known as a duct-type exhaust high-efficiency filter device. Since the filter has retained highly active or highly toxic harmful aerosols during the working process, it is necessary to ensure that the filter does not have any contact with the external environment during the replacement process. 
  The entire process of filter replacement is carried out in a sealed bag, so it is called bag entry Bag out the filter. Its use can effectively prevent the spread of harmful aerosols and avoid biological hazards to personnel and the environment. It is a filter device used in specific biological risk environments to remove harmful biological aerosols in the exhaust air. Generally, it has in-situ disinfection. And leak detection function, in the process of use, it needs to be equipped with biological sealing valve and other components.
  Commonly used functional units include: Pre-filter, HEPA, HEGA, Precision ScanTest Sections, etc. The simplest filter unit only contains a high-efficiency particle filter. If there is no other filtering device before the air enters the bag into the bag and exits the filter, you need to configure a pre-filter to protect the high-efficiency particle filter. If the filtered gas has adsorption requirements, it is equipped with high-efficiency Gas adsorption filter; detection unit refers to automatic (or manual) fine scanning detection device.
Advantages
1: High Security: The AC.Safe. BIBO filter equipment has air tight welding housing to prevent the leakage of contaminated material. In addition, A differential gauge with screen and output port is standardly equipped on the bag in-bag out filter box to test the leakage. 

2: Negative pressure design: The AC.Safe. filtering system unit could sustain -5000Pa pressure. 

3: Door easily remove:The door are specially designed for the tightness and convenience. Only four bolt fastening could easily remove the access doors, but the inner sealing also protect the leakage.

4: Two HEPA filters equipped: the first HEPA is the mainly filtering step, and second HEPA is the protective HEPA.
How to change the HEPA filter and F9 Filters?
Step1: Open the doors

Step 2: Tight the safety bag

Step 3: Pull out of the HEPA

Step 4: seal Bag/Remove

Step 5: Tight another bag with Hepa, remove the first bag.

Step6: Insert the Hepa and close the door

2 Stage HEPA+F9 Filter Housing Technical Data

Model

Airflow

(m3/h)

HEPA Size

(mm)

HEPA Quantity (pcs)

F9 Filter Size

(mm)

F9 Filter Quantity

(pcs)

BIBO-FH-500CFM

850

484*484*220

1

484*484*46

1

BIBO-FH-1000CFM

1700

592*592*292

1

592*592*46

1

BIBO-FH-2000CFM

3400

592*950*292

1

592*950*46

1

BIBO-FH-4000CFM

6800

592*950*292

2

592*950*46

2

BIBO-FH-6000CFM

10200

592*950*292

3

592*950*46

3

Video
Please send your message to us
*Email
Tel
*Title
*Content
Upload
  • Only supports .rar/.zip/.jpg/.png/.gif/.doc/.xls/.pdf, maximum 20MB.
subscription
KEEP IN TOUCH
Please send your message to us
*Email
Tel
*Title
*Content
Upload
  • Only supports .rar/.zip/.jpg/.png/.gif/.doc/.xls/.pdf, maximum 20MB.